M3500B/M5500B Service Manual
3
-
17
Momentary Messages
3
Audio Tones
The M3500B/M5500B emits tones to alert you to its status.
No Data Card Present
A Data Card is not in the M3500B/
M5500B.
Turn the M3500B/M5500B off and insert a
Data Card prior to the first event for the
patient.
Key Inactive
The key pressed is currently inactive (i.e.
and
are inactive in
AED Mode).
Use the appropriate mode for the key.
Stop Pacer
is pressed while pacing pulses
are being delivered.
Stop pacing before changing the pacing
mode.
Table 3-4 Momentary Messages (Continued)
Message
Possible Cause
Corrective Action
Sync
Pace
Mode
Table 3-5 Audio Tones
Tone
Meaning
Suggested Action
At power on, a low tone of approx. 1
second followed by a series of
higher pitched, short tones.
Normal power on sequence.
None needed.
At power on, a continuous beep of
about 7 seconds.
System failure - processors not
communicating.
Turn power off, then on. If problem
reoccurs, replace Control PCA.
At any time, a repeating tone that
alternates between two frequencies.
The unit emits this tone beginning 1
minute before shutdown due to low
battery charge.
1. Connect a Power Module or
replace the battery with one that
is fully charged.
2. Replace Power Module.
3. Service the unit. See "Trouble-
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...