Operational Problems
3
-
30
Troubleshooting
Figure 3-3 Power Module Indicators
Power Indicator
Charge Done
Indicator
Charging Indicator
Table 3-16 Power Module Indicator Matrix
INDICATORS
SYSTEM STATE
MEANING
POWER
CHARG-
ING
DONE
Con-
nected
to Mains
(AC/DC) ?
Con-
nected
M3500B/
M5500B ?
Battery in
Place?
Unit Power
ON ?
R = RED G = GREEN
OR = ORANGE/AMBER
Y = YES N = NO --- = don’t care
OFF
OFF
OFF
N
---
---
---
Normal
G
OFF
OFF
Y
---
---
---
Normal
G
OFF
OFF
Y
Y
N
---
Normal
G
OR
OFF
Y
Y
Y
---
Battery is charging.
G
G
OFF
Y
Y
Y
---
Battery 90% charged.
G
G
G
Y
Y
Y
---
Battery fully charged.
R
OFF
OFF
Y
Y
---
---
If Power light was green before
connecting to unit, fault in
M3500B/M5500B.
R
OFF
OFF
Y
N
---
---
Mains voltage low or
Fault in Power Module.