ITS Uninterruptible Power Supply Test and Inspection Procedures
Table 5-9 lists which tests you must perform when performing service events on
IntelliVue Telemetry System UPS.
Table 5-9: ITS UPS Test and Inspection Requirements
When Performing this
Service Event...
Complete These Required Test(s)
Installation
Visual, Power On
Preventive Maintenance
N/A
Any component repair or replacement
Power On
Hardware Upgrade
N/A
Software Upgrade
N/A
All other Service Events
Visual, Power On
Table 5-10: ITS UPS Test and Inspection Matrix
Test Block
Name
Test or Inspection to Perform
Expected Results
Information to
Record on
Service Record
UPS Visual
Inspect all system components for
obvious damage. Visually inspect all
system components, cables, and
connectors. Check for signs of abrasion,
wear, or other damage.
No visible damage
V:P or V:F
where:
P = Pass
F = Fail
UPS
Power On
With power connected to each active
Network device, observe that all lights
visible on the front panel are in proper
status and that no error conditions are
shown.
Devices power up
into expected
status; no error
indications are
shown.
PO:P or PO:F
where:
P = Pass
F = Fail
UPS
Performance
No performance check is required.
P:NA
where:
NA=Not
Applicable
UPS
Revision Check
No revision check is required.
RC:NA
where:
NA=Not
Applicable
UPS
Safety
No safety test is required
S:NA
where:
NA=Not
Applicable
Summary of Contents for IntelliVue Telemetry System Infrastructure
Page 8: ...viii Contents ...
Page 14: ...xiv About This Guide ...
Page 100: ...3 22 Chapter 3 Installing and Configuring the ITS Figure 3 10 APC Filter Configuration Screen ...
Page 162: ...5 12 Chapter 5 Troubleshooting and Testing ...
Page 168: ...A 6 Chapter A Installing Multiple ITSs at a Single Hospital Site ...
Page 180: ...Index 8 Index ...