IntelliVue Telemetry System Specifications
IntelliVue Telemetry System Infrastructure Installation and Service Guide
1
-
25
Table 1-13: UPS Specifications
UPS
Requirements
Parameter Specification
Input Voltage
1.4 GHz ITS
2.4 GHz ITS
/-10%
100, 120, 127, 220, 240 VAC ± 10%
Input
Frequency
60Hz +/-3Hz minimum
50, 60Hz +/-3Hz
UPS AC
Input Port
100, 120VAC model: one NEMA 5-15P plug.
230 VAC model: one IEC 320 C14 plug.
Number of UPS
AC output ports
100, 120VAC model: 3 minimum (NEMA 5-15R output receptacles),
desirable to have 5 ports or greater.
230V model: 3 minimum (IEC 320 C13 output receptacles), desirable
to have 5 ports or greater.
Power delivery
UPS shall provide at least 225Watts for > 90 seconds to power
12 APs, 1 Sync Unit, 2 APC, 1 POE, & 1 Switch from one UPS unit with 5
UPS ports.
Or:
12 APs, 1 Sync Unit, 1 APC, 1 POE, from one UPS unit with 3 UPS ports.
Additional UPS would be needed to power redundant APC & Switch.
Power Fail
Transition
UPS must be responsive to power fail conditions such that ITS system
does not experience reboot due to transition to temporary battery
backup power.
Power Fail Alert
UPS shall provide local auditory response to power fail conditions.
EMI Filtering
Provide AC line EMI filtering over 100KHz to 30MHz on UPS ports.
EMC
FCC part 15 class A or B certification, CE Mark: CISPR 22/ EN 55022,
and CISPR 24 / EN 55024.
Safety
UL certification, EN 62040 (Low Voltage Directive).
Mounting
Rack Mounting is desirable.
Summary of Contents for IntelliVue Telemetry System Infrastructure
Page 8: ...viii Contents ...
Page 14: ...xiv About This Guide ...
Page 100: ...3 22 Chapter 3 Installing and Configuring the ITS Figure 3 10 APC Filter Configuration Screen ...
Page 162: ...5 12 Chapter 5 Troubleshooting and Testing ...
Page 168: ...A 6 Chapter A Installing Multiple ITSs at a Single Hospital Site ...
Page 180: ...Index 8 Index ...