The Test Signals Function
3-60 Setting up your Monitor
S
e
tt
ing up y
our
M
onito
r
System Self-Test Values:
Module
Test Numeric Limits
Test Waveforms
ECG and ECG/
RESP
ECG
100 bpm in ADULT mode
125 bpm in NEO/PEDI mode
Simulated square wave and numeric
RESP
15 rpm in ADULT mode
30 rpm in PEDI mode
55 rpm in NEO mode
Simulated Resp wave and numeric
Pressure
Simulated square wave and numerics
Systolic
120 mmHg ADULT
60 mmHg PEDI/NEO
Diastolic
0 mmHg ADULT/PEDI/NEO
SpO
2
/Pleth
SpO
2
100%
Numeric only
Pleth
Wave on display
Simulated Wave
tcpO
2
/tcpCO
2
Numeric only
tcpO
2
60 mmHg (8.0 kPa)
tcpCO
2
40 mmHg (5.3 kPa)
CO
2
Simulated square wave and numerics
ETCO
2
40 mmHg (6.0 kPa)
IMCO
2
0 mmHg (0.0 kPa)
AWRR
25 rpm
Temperature
40°C (104°F)
Numeric only
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...