Qualifying indicators (which have specific meaning) appear in the Blood
Review task window. The following list identifies and defines all
indicators used to qualify blood analysis results in the Blood Review task
window.
•
A “T” next to the time indicates that the data was transferred to the
monitor using the Data Transfer module.
•
A question mark beside a time (?7:15) indicates the blood
measurement occurred prior to clock changes at the bedside or
central station.
•
A question mark preceding a blood measurement value (?5.0)
indicates that blood measurement has been invalidated by a user.
•
A question mark standing alone in a blood measurement field (?)
indicates that no valid result could be derived from the sensor
signals.
•
The ' symbol at the beginning of the blood measurement label
('BEecf) indicates the measurement has been calculated by the
blood analysis module.
•
The “&” symbol before a label (e.g. &PCO2) indicates the
measurement has been adjusted to the patient's temperature.
•
The > symbol preceding a blood measurement value (>3.0) indicates
the value is greater than the upper limit of the reportable range.
•
The < symbol preceding a blood measurement value (<3.0) indicates
the value is less than the lower limit of the reportable range.
If configured to “Enabled” in the Blood Analysis configuration task
window, O2*, the Free Fields (Field1, Field2, and Field3), and Operator ID
of the person performing the blood measurement appear after the last
measurement (in the order listed).
Viewing Blood
Review
Reference
Ranges
You can compare an adult arterial/venous blood result value against a
corresponding adult arterial/venous blood reference range. The blood
results on the Blood Review task window can be within the reference
range or outside the reference range. You can view reference ranges by
selecting the
key on the Blood Review task window
display. The reference ranges correspond to the rightmost time displayed
in the Blood Review task window. The reference range information that
appears corresponds to the last measurement. The out of range indication
refers to the reference range at the time of the measurement.
Ranges/Units
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...