Power
Source
Require-
ments
The CMS systems and the Anesthetic Gas Module can be operated from
an AC source of 100 - 120V
±
10%, or 220 - 240V
±
10%,
50 - 60Hz.
Typical input currents for the monochrome and color systems are:
Typical input current for the Anesthetic Gas Module is:
100 - 240V = 1.0 amps
Typical input current for the V24 and V26 is:
90 - 250V = 150VA
Grounding
the System
To protect the patient and hospital personnel, the cabinet of the installed
equipment has to be grounded. The equipment is supplied with a
detachable 3-wire cable which grounds the instrument to the power line
ground (protective earth) when plugged into an appropriate 3-wire
receptacle. If a 3-wire receptacle is not available, consult the hospital
electrician.
Warning
Warning
Do not use a 3-wire to 2-wire adapter.
Monochrome CRT
100 - 120V = 2.2 amps
200 - 240V = 1.2 amps
Color CRT
100 - 120V = 2.6 amps
200 - 240V = 1.4 amps
Color Flatscreen
110 - 120V = 2.0 amps
200 - 240V = 1.0 amps
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...