Maintenance Checks
Monitor Installation and Patient Safety 11-15
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
This recommendation does not supersede any local procedures or legal
requirements.
Electrical Safety Tests
At least every two years, after
system installation, after any
repairs, or as needed.
See “Testing and Maintaining the
CMS” in the “CMS Service Guide”
or “Testing and Maintaining the
V24 and V26“in the “V24 and V26
Service Guide”.
Testing of all cables and cords.
Daily.
See “Testing and Maintaining the
CMS” in the “CMS Service Guide”
or “Testing and Maintaining the
V24 and V26“in the “V24 and V26
Service Guide”.
Calibration of the measurement
modules where appropriate
(For EU/EFTA countries only)
Every 2 years.
See individual module chapters in
Volume 2 of this manual for
details.
User Tests
Module dependent. Please refer to
chapter 13 “Maintenance” of this
manual or chapter 3 of the CMS
Service Guide “Maintaining the
CMS” for details.
See chapter 13 “Maintenance” of
this manual for details.
Maintenance
Frequency
Source of Information
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...