Maintenance Checks
11-16 Monitor Installation and Patient Safety
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
Patient
Cables and
Leads
Inspect the patient cables and leads and their strain reliefs for general
condition. Examine cables carefully to detect breaks in the insulation
and to ensure that they are gripped securely in the connectors of each
end to prevent rotation or other strain.
To ensure operation of the Anesthetic Gas Module within specified
limits:
•
calibrate the Anesthetic Gas Module at least once every 12
months, or whenever the validity of the readings is in doubt.
•
check the ventilator fan in the Anesthetic Gas Module for proper
operation every 6 months.
All checks which require the instrument to be opened must be made by
qualified hospital service personnel. Safety and maintenance checks can
also be made by Philips personnel.
Warning
Warning
Failure on the part of the responsible individual hospital or
institution employing the use of this equipment to implement a
satisfactory maintenance schedule may cause undue equipment
failure and possible health hazards.
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...