Controls and Connectors
11-22 Monitor Installation and Patient Safety
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
Note—
Do not apply a voltage of more than
±
12V to any of these outputs.
The connectors on the rear panel of the computer module (together with
their connecting function cards) are:
1. System Power Connector; on the M1046A Computer Module, this
is a connector used to input the 60V dc power supply from the
CRT display module. It is not for use on the M1046B Computer
Module but is instead fitted with a stub connector.
2. Equipotential Grounding Terminal; this is used to connect the
computer module to the hospital´s grounding system.
3. Philips monitoring network connectors (Philips Network
Interface); these are Philips monitoring network connectors, used
to input/output information to/from a Philips monitoring network.
4. Human Interface Connector (Utility CPU); this is a Philips HIL
connector, used to input information from the keypad in the
display module.
5. Nurse Paging Relay output; this is a mini-phone jack, used to
connect to the hospital´s Nurse Paging System.
6. Analog Interface Connector; a 37-pin “D” type connector is used to
output information to analog devices.
7. RS232 Connectors (RS232 Interface); these are 25-pin “D” type
connectors, used to output information to printers or computers,
RS232 Connectors
±
12 V
HDLC Connector
5 V
Video Out (9)
5 V
Flatscreen Remote Power Supply On/Off
15 V
Video Out (11)
60 V
Maximum Voltages
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...