![background image](http://html.mh-extra.com/html/philips/m1165/m1165_users-reference-manual_303827381.webp)
Performance Assurance Checks
Maintenance 13-15
M
a
int
e
na
nc
e
Performing
the ECG
Module and
ECG/RESP
Self-Test
Auto Check
- None
Self-Test
The ECG and ECG/RESP self-test may be performed at any
time as the lead selector and pace pulse detection are inhibited during the
check. The module will produce ECG test signals at the beginning of the
processing circuitry, and process the test signals in the same way patient
signals are processed. The test signals last about 30 seconds and then the
display returns to the normal monitoring mode.
TEST - Press
.
RESULT - If the module is functioning correctly, the following heart rate
and respiration are displayed without the alarms being activated:
ECG 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode
RESP 15 rpm in ADULT mode 30 rpm in PEDI mode 55 rpm in NEO mode
Cardiac Output
5 l/min (+0.1 l/min)
(Blood Temp: 37°C)
(Inj Temp: 0°C)
No wave
tcpO
2
/tcpCO
2
No wave
tcpO
2
60 mmHg (8.0 kPa)
tcpCO
2
40 mmHg (5.3 kPa)
CO
2
Simulated square wave
ETCO
2
40 mmHg (6.0 kPa)
IMCO
2
0 mmHg (0.0 kPa)
AWRR
25 rpm
Temperature
40°C (104°F)
No wave
Module
Test Numeric
Test Waveform
Monitor Setup
Test Signals
Pts
Refernce
Signals
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...