![background image](http://html.mh-extra.com/html/philips/m1165/m1165_users-reference-manual_303827061.webp)
Using an ITE Display
The CMS and V24 and V26 Patient Monitors 1-39
EMC
1. The display must either fulfill the requirements of the EN60601-1-2
{IEC 601-1-2}, or
it must conform to the requirements of the EMC standard for ITE
devices, EN50082-1: 1997 and EN50081-1 / CISPR 22.
2. The video cable between M1167A/77A CMS Patient Monitoring
System and the ITE display must not exceed a length of 3.0m. For
cables exceeding 3.0m EMC testing according to IEC 801-4:1988 has
to be done.
Perfor-
mance
Require-
ments
Philips cannot assure compliance with the ANSI/AAMI EC-13 Standard for
Cardiac Monitors, Heart-rate Meters and Alarms when using ITE displays.
Compliance with the ECG aspect ration and 25 mm/s
±
10% sweep-speed
specifications can only be assured when using the Philips Displays
M1092A, M1094B and M1095A. In addition, the following display
specifications are recommended:
Specification
Requirement or Value
Units
Resolution (pixel addressability)
1024 x 768
Dots x lines
Vertical Refresh Rate
60
Hz (non-interlaced)
Red, Green and Blue Video Inputs
~0.7
V p-p
Vertical & Horizontal Multi-Sync Inputs
5
V TTL
Video-Cable Connector
HD15male
Summary of Contents for M1165
Page 12: ...Responsibility of the Manufacturer xii ...
Page 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Page 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Page 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Page 228: ...Loading Paper 6 56 Recording Functions ...
Page 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Page 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Page 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Page 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Page 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...