Instrument Test
B-2
Specifications
,QVWUXPHQW7HVW
A self-test may be started by pressing and holding both the
and
keys while turning on the cardiograph with the
key. The instrument performs an internal self-test. The test results are
printed on the report for use by service personnel. This self-test runs
continuously until the cardiograph is turned to Standby (off).
3DWLHQW6DIHW\
Patient Isolation: Less than 20 µA leakage with 120 VAC, 60 Hz or less
than 50 µA leakage with 240 VAC, 50 Hz with patient cable.
Defibrillation Protection: Protected against damage from 400 joule
defibrillator discharges.
3RZHUDQG(QYLURQPHQW
Line Power: 100 to 120 VAC, 50 to 60 Hz (at 115 VAC power-switch
setting); 220 to 240 VAC, 50 to 60 Hz (at 230 VAC power-switch set-
ting); 50 VA Maximum.
Fuse: 0.4 A for both the 115 and 230 VAC switch settings.
Environmental Operating Conditions: 10 to 40
°
C (50 to 104
°
F), 15 to
80% relative humidity, non-condensing, up to 4,550 m (15,000 ft.) alti-
tude.
Environmental Storage Conditions: 0 to 50
°
C (32 to 122
°
F), 15 to 90%
relative humidity, non-condensing, up to 4,550 m (15,000 ft.) altitude.
Cardiograph Dimensions: 43 by 39 by 10 cm (17.0 by 15.3 by 3.9 in.).
Cardiograph Weight: 8.5 kg (19.8 lbs.).
$XWR
0DQXDO
2Q6WDQGE\
Summary of Contents for M1772A
Page 1: ...I N S T R U C T I O N S F O R U S E C a r d i o g r a p h PageWriter 100 M1772A ...
Page 2: ......
Page 3: ...PageWriter 100 M1772A Cardiograph 3DJH ULWHU 0 DUGLRJUDSK ...
Page 10: ...Contents 4 ...
Page 22: ...AC and DC Battery Operation 1 12 Getting Acquainted ...
Page 38: ...Choosing a Report Format 2 16 Recording an ECG LJXUH 0DQXDO HDG ...
Page 44: ...Identifying ECG Problems 3 6 Troubleshooting ...
Page 60: ...Calling for Service 4 16 Maintaining the Cardiograph Spain Tel 34 91 631 31 00 ...
Page 76: ...Glossary Glossary 4 ...
Page 80: ...Index 4 ...
Page 81: ......