no connection with the tEns units
- This ‘no connection’ symbol on the screen
indicates that there is no connection between the
remote control and the TENS units. This may have
several causes:
1 The rechargeable battery of the TENS units is
empty.
2 The TENS units are switched off.
3 The remote control is out of reach of the TENS
units.
4 If neither of the above applies, there is something
wrong with the TENS units.
channel interrupted
- The ‘channel interrupted’ symbol (the cross)
on the screen indicates that the corresponding
channel is interrupted.
- If a channel is interrupted, this may have the
following causes:
1 There is no contact between the TENS units and
the electrodes.
2 The electrodes are not attached to the skin
properly.
- When you have made sure the electrodes are
properly attached to the skin and the TENS units
are properly connected to the electrodes, press
the + side of the intensity button to check if the
channel is properly connected. If it is properly
connected, the cross disappears and the intensity
value appears.