indicator signals on the tEns units
Battery status indicator
- The battery status indicator on the TENS units
flashes:
1 The battery status indicator flashes orange
briefly when the rechargeable battery is empty
and then the TENS units switch off.
2 The battery status indicator flashes green when
the TENS units are charging.
- The battery status indicator is burning solid:
1 The battery status indicator is solid orange when
the rechargeable battery of the TENS units is
low. The rechargeable battery has only 30% of
its total capacity left. The battery status indicator
also turns solid orange when the batteries of the
remote control are low.
2 The battery status indicator is solid green when
the rechargeable battery is fully charged.
treatment status indicator
- The treatment status indicator on the TENS units
flashes:
1 The treatment status indicator flashes green
when the TENS units are switched on and in
standby.
2 The treatment status indicator flashes yellow
during treatment to indicate that the treatment
is in progress.
3 The treatment status indicator flashes red when
the TENS units overheated. When the TENS
units become overheated, the treatment stops
automatically and the ‘no connection’ symbol
appears on the display. Detach the TENS units
from the electrodes and let them cool down.
As soon as they have cooled down sufficiently,
reattach them and resume treatment.
English