Chapter 6
Reports and Software Downloads
1049766 Rev A
V60 Ventilator Service Manual
6-21
6.4 Downloading
Ventilator Software
Use the Respi-Link Remote Diagnostic System to download V60 ventilator
software. If your institution is not Respi-Link capable, contact Respironics to
request Respi-Link information from your local Field Service Specialist or
Regional Service Provider.
If your institution is Respi-Link capable, contact Respironics to have a
software package deployed. See Appendix A for installation instructions using
Respi-Link.
NOTE:
Note: If you have replaced the CPU PCBA, please contact Respironics to
have your local Field Service Specialist or Regional Service Provider
install the software for you.
For technical support and customer service, contact:
USA and Canada: 800-345-6443 or 724-387-4000
Respironics Europe, Africa, Middle East: +33-1-47-52-30-00
Respironics Asia Pacific: +852-3194-2280
6.5 Programming the
Ventilator Serial
Number and Power-
On Hours
The ventilator serial number and power-on hours must be re-programmed to
the CPU PCBA whenever you replace the CPU PCBA.
CAUTION:
Only qualified service technicians are to program the ventilator serial
number and power-on hours.
NOTE:
V60 ventilator software must be installed before programming the
ventilator serial number or power-on hours.
6.5.1
Programming the Ventilator Serial Number
NOTE:
Do
not
connect the null modem cable until instructed to do so.
Follow these steps to re-program the ventilator serial number to a new CPU
PCBA:
1. Enter Diagnostic mode by pressing and holding the nav-ring
Enter
button and turn on the ventilator by pressing
On/Shutdown
. Within 5
seconds, press the nav-ring
Enter
button again to enter Diagnostic
mode.
2. Turn on the PC and connect a standard 9-pin male/female RS 232
null modem cable and a 9-pin to 25-pin female/male adapter between
the PC serial port and the V60 ventilator.
3. Click on the HyperTerminal or Tera Term icon to launch the program
that enables communication with the V60 ventilator.