Chapter 10
10-6
V60 Ventilator Service Manual
1049766 Rev A
Specifications
10.7 Environmental
specifications
10.8 Pneumatic
specifications
Table 10-8: Environmental specifications
Parameter
Specification
Temperature
Operating: 5 to 40
o
C (41 to 104
o
F)
Storage: -20 to 50
o
C (-4 to 122
o
F)
Relative humidity
Operating: 15 to 95% (noncondensing)
Storage: 10 to 95% (noncondensing)
Barometric pressure
600 to 765 mmHg (approximately -51 to 1951 m/
-200 to 6400 ft relative to sea level)
Table 10-9: Pneumatic specifications
Parameter
Specification
High-pressure oxygen supply
Pressure: 2.76 to 6.00 bar / 276 to 600 kPa / 40 to
87 psig
Flow: 175 SLPM
Connector: DISS male, DISS female, NIST, SIS
Air supply
Integrated blower
Inspiratory outlet (To patient
port)
Connector: ISO 15 mm female/22 mm male conical