background image

12

P

ROBLEM

W

HY

 

IT

 H

APPENED

W

HAT

 

TO

 D

O

The therapy device is 
operating but the humidifier’s 
airflow is low or stopped.

The humidifier has an 
airflow obstruction.

Contact your home care provider. Return the humidifier and power 
supply to your provider to determine the problem.  

The water tank’s middle seal 
is damaged.

The seal may become 
damaged during use.

Check the seal for any tears or other damage. If it is damaged, 
contact your home care provider.  

Excessive condensation in 
the tubing.

The heater plate 
temperature setting is too 
high.

Reduce the heater control setting.

The humidifier is 
positioned incorrectly.

Verify that the humidifier and therapy device are away from air 
conditioning equipment.

The humidifier door has 
been accidentally removed 
from the humidifier.

Humidifier door was lifted 
past the locked open 
position.

The humidifier door can be easily reattached. 

With the water tank removed, align the hinge side of the door with 
the back of the humidifier in a fully open position (see figure below). 
Make sure the hinges are aligned with the grooves of the humidifier. 
Rotate the door closed. You will need to apply pressure until the 
door snaps past the locked open position. Verify that the door is 
working properly.

If further assistance is needed, contact your home care provider.

The humidifier has fallen off 
your table or night stand.

The humidifier may not 
have been properly seated 
on the night stand, or the 
placement of the tubing 
may have caused the device 
to fall. 

Always make sure your humidifier is placed on a hard, flat surface 
so the rubber feet on the bottom of the humidifier base can adhere 
to the surface (make sure there is no fabric under the base). The 
humidifier must be level for proper operation.

Also, place the humidifier away from the edge of the night stand or 
table, so it doesn’t accidentally get knocked off the table.

If the humidifier falls and water gets into the therapy device, drain all 
water out of the therapy device. Allow it to air dry to make sure it is 
completely dry before reapplying power.

If the placement of the tubing causes the humidifier to fall, make sure 
that you use proper hose management when setting up your device. 
Route the tubing behind the bed’s headboard.

If the humidifier does not operate correctly after falling, contact your 
home care provider or Respironics. 

 Note:

  For information on troubleshooting your therapy device, see the manual included with your therapy device.

Summary of Contents for respironics

Page 1: ...System One Heated Humidifier USER MANUAL ...

Page 2: ... 2010 Koninklijke Philips Electronics N V All rights reserved ...

Page 3: ...erapy Device 4 Disconnecting the Therapy Device 5 Daily Use 5 Home Cleaning Water Tank 9 Home Cleaning Humidifier Base 9 Hospital and Institution Cleaning 10 Service 10 Traveling with the System 10 Troubleshooting 11 Specifications 13 Disposal 14 How to Contact Respironics 14 EMC Information 15 Limited Warranty Back Page ...

Page 4: ...py device Empty and clean the water tank daily to prevent mold and bacteria growth Wipe the seal completely Note Please see the Limited Warranty section of this manual for information on warranty coverage Cautions A caution indicates the possibility of damage to the device Do not place the humidifier directly onto carpet fabric or other flammable materials Do not place the device in or on any contai...

Page 5: ...t Port Connects to the outlet port on the therapy device Air Outlet Port conical 22 mm Connect the patient circuit here Heater Plate Warms the water in the water tank Humidifier Door Open the door to access the water tank Humidifier Door Lever Lift up on the lever to open the humidifier door Water Tank The removable water tank holds the water for humidification Maximum Fill Lines The fill lines indicat...

Page 6: ...he side cover and pull the cover away from the unit Therapy Device Side Cover Tab 3 Next line up the components side by side Make sure that the guides on the humidifier fit into the slots on the therapy device and the air outlet port on the therapy device fits into the air inlet port on the humidifier 4 Simply press the two units together until they snap into place Make sure that the therapy device an...

Page 7: ...AUTION Do not place the device in or on any container that can collect or hold water Take precautions to protect furniture from water damage CAUTION Do not turn the humidifier on without the water tank installed The Humidifier number setting must remain on 0 if there is no water in the water tank CAUTION Do not attempt to fill the tank while it is still inside the humidifier WARNING Allow the humidifie...

Page 8: ...d on the front and both sides of the tank Tank Base Tank Lid MaximumWater Fill Lines Note Clean the water tank before first use Refer the the Home Cleaning section in this manual CAUTION Use only room temperature distilled water in the tank Do not put any chemicals or additives into the water Possible airway irritation or damage to the water tank may result CAUTION Do not fill the water tank above t...

Page 9: ...ter to spill into the humidifier or therapy device CAUTION Do not move the humidifier while the water tank has water in it 7 Plug the power supply cord into the back of the therapy device 8 Plug the AC power cord into the power supply provided with your therapy device and then plug the AC power cord into an electrical outlet that is not controlled by a wall switch 9 Attach the flexible tubing from th...

Page 10: ... without the water tank installed The Humidifier number setting must remain on 0 if there is no water in the water tank IMPORTANT The ideal humidity setting depends on room temperature and humidity Initially a setting of 2 is recommended You can adjust this setting at any time Turn the Wheel to change the setting Note The Humidifier Icon will light up when therapy is active and the heater plate is o...

Page 11: ...Before using the tank fill it with distilled water per the directions in the Daily Use section no higher than the maximum fill line located on the front and sides of the tank 7 Reassemble the tank by placing the hinges on the tank lid over the 2 tabs on the back of the tank base Close the lid until the tab on the lid snaps back under the lip in the tank base Inspect the tank When it is closed correc...

Page 12: ...ece Discard and replace any damaged parts Service The humidifier does not require routine servicing If any part of the humidifier is worn or damaged contact Respironics or your home care provider See the Troubleshooting section later in the manual for additional information Traveling with the System Packing the System 1 Remove the water tank and empty all water 2 Put the empty water tank back into t...

Page 13: ...ly seated in the humidifier Remove the water tank from the humidifier base and then slide the tank back in the humidifier making sure it is pushed back as far as it can go If high leak persists issue may be caused by a misaligned or damaged seal Seals are not user serviceable components Please contact your home care provider or Respironics The humidifier LED icon on the therapy device is flashing There...

Page 14: ...apply pressure until the door snaps past the locked open position Verify that the door is working properly If further assistance is needed contact your home care provider The humidifier has fallen off your table or night stand The humidifier may not have been properly seated on the night stand or the placement of the tubing may have caused the device to fall Always make sure your humidifier is placed...

Page 15: ...ctrical Equipment EN ISO 8185 2007 General Requirements for Humidification Systems Electrical When the heated humidifier is used with a Respironics therapy device AC Power Consumption 100 240 VAC 50 60 Hz 2 1 A DC Power Consumption 12 VDC 5 0 A Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Shock Type BF Applied Part Degree of Protection against In...

Page 16: ...tact Respironics To have your device serviced contact your home care provider If you need to contact Respironics directly call the Respironics Customer Service department at 1 724 387 4000 or 49 8152 93060 You can also use the following addresses Respironics Inc 1001 Murry Ridge Lane Murrysville PA 15668 Respironics Deutschland Gewerbestrasse 17 82211 Herrsching Germany ...

Page 17: ...rge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical Fast Transient Burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for supply mains 1 kV for input output lines Mains power quality should be that of a typica...

Page 18: ...ss the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocati...

Page 19: ...pironics Inc products and reserves the right to charge dealers for warranty service of failed product not purchased directly from Respironics or authorized distributors Respironics Inc disclaims all liability for economic loss loss of profits overhead or consequential damages which may be claimed to arise from any sale or use of this product Some states do not allow the exclusion or limitation of i...

Page 20: ...1058289 1057396 Rev 01 JR 7 28 2010 EN INTL ...

Reviews: