Specifications
Environmental
Operating Temperature: 5° to 35° C (41° to 95° F)
Storage Temperature: -20° to 60° C (-4° to 140° F)
Relative Humidity (operating & storage): 15 to 95% (non-condensing)
Atmospheric Pressure: 77 to 101 kPa (0 - 2286 m / 0 - 7500 ft)
Physical
Dimensions: 18 x 14 x 10 cm (7” L x 5.5” W x 4” H)
Weight: Approximately 0.89 kg (1.95 lbs.)
Water Capacity
325 ml (11 oz.) at recommended water level
Standards Compliance
This device is designed to conform to the following standards:
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment
EN ISO 8185:2007 General Requirements for Humidification Systems
Electrical
(When the heated humidifier is used with a Respironics therapy device)
AC Power Consumption: 100 – 240 VAC, 50/60 Hz, 2.1 A
DC Power Consumption: 12 VDC, 5.0 A
Type of Protection Against Electric Shock: Class II Equipment
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Protection against Ingress of Water: Drip Proof, IPX1
Mode of Operation: Continuous
Electromagnetic Compatibility: The device meets the requirements of EN 60601-1-2, 2nd edition.
Heater Plate
Max Temperature: 75° C (167° F)
Pressure Drop with Humidifier
Max.: 0.3 cm H
2
O at 60 LPM flow
Humidity
Humidity
min
Output: 10 mg H
2
O/L
Measured @ max flow, 35° C, 15% RH.