Intended Use
The Heated Humidifier is an accessory for the Respironics REMstar and BiPAP therapy devices to provide moisture
to the patient circuit. It is intended for use in spontaneously breathing patients weighing over 30 kg (66 lbs), in the
home or hospital/institutional environment, who use mask-applied positive pressure ventilation therapy.
Warnings
A warning indicates the possibility for injury to the user or the operator.
• Use the humidifier only for its intended use as described in this manual. Use only accessories recommended by Respironics.
• Periodically inspect the humidifier for signs of wear or damage. Never operate the humidifier if any parts are damaged, if it is not
working properly, or if the humidifier has been dropped or mishandled. Do not use the humidifier if the water tank is leaking or
damaged in any way. Have any damaged parts replaced before continuing use.
• Periodically inspect the humidifier power cord for signs of wear or damage. If it becomes worn or damaged, contact Respironics
or your home care provider for a replacement.
• The humidifier must always be positioned below the breathing circuit connection at the mask and the air outlet on the device. The
humidifier must be level for proper operation.
• Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank. A burn
may result from: touching the heater plate, coming in contact with the heated water, or touching the tank pan.
• This equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
• When installing the water tank, do not allow any water to spill into the humidifier or therapy device.
• If you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, if it has been
dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, disconnect the power cord from the
therapy device and discontinue use. Contact your home care provider.
• Before cleaning the humidifier, always remove from the therapy device.
• Empty and clean the water tank daily to prevent mold and bacteria growth. Wipe the seal completely.
Note:
Please see the “Limited Warranty” section of this manual for information on warranty coverage.
Cautions
A caution indicates the possibility of damage to the device.
• Do not place the humidifier directly onto carpet, fabric, or other flammable materials.
• Do not place the device in or on any container that can collect or hold water. Take precautions to protect furniture from water damage.
• Do not fill the water tank above the maximum fill line. Damage to the humidifier or therapy device may occur. If the water tank is
overfilled, water may leak out of the tank inlet (located on the back of the tank) when installing the tank lid.
• Use only room temperature distilled water in the tank. Do not put any chemicals or additives into the water. Possible airway
irritation or damage to the water tank may result.
• Remove the tank, empty all water, and replace the empty tank before transporting the humidifier base.
• Do not attempt to fill the water tank while it is still inside the humidifier.
• To avoid spilling, do not disconnect the humidifier from the therapy device with water in the tank. Remove the water tank from
the humidifier before disconnecting the therapy device.
• Do not turn the humidifier on without the water tank installed. The Humidifier number setting must remain on 0 if there is no
water in the water tank.
• The humidifier door must be set in the open position before removing the water tank. Do not remove the water tank without
making sure that the humidifier door locks into the open position.
• Do not move the humidifier while the water tank has water in it.
• Use a mild liquid dishwashing detergent only for either hand washing or when using a dishwasher.
• Only the hospital and institution cleaning procedures listed in this manual are recommended by Respironics. Use of other cleaning
and disinfecting processes, not specified by Respironics, may affect the performance of the product.
Symbol Key
The following symbols appear on the device:
S
YMBOL
D
EFINITION
S
YMBOL
D
EFINITION
Maximum Fill Line
Drip Proof Equipment
Hot Water, Hot Surface
Type BF Applied Part
Follow Instructions for Use
Separate collection for electrical and electronic
equipment per EC Directive 2002/96/EC.