Instructions for Use | Alarms and System Messages
Alarm and Event Log
63
E
T
CO
2
1.
Set the High EtCO
2
alarm limit below the measured EtCO
2
value.
2.
Confirm that the High EtCO
2
alarm activates.
3.
Restore the alarm limit.
4.
Set the Low EtCO
2
alarm limit above the measured EtCO
2
value.
5.
Confirm the Low EtCO
2
alarm activates.
6.
Restore the alarm limit.
P
ULSE
R
ATE
1.
Set the High Pulse Rate alarm limit below the measured Pulse Rate value.
2.
Confirm that the High Pulse Rate alarm activates.
3.
Restore the alarm limit.
4.
Set the Low Pulse Rate alarm limit above the measured Pulse Rate value.
5.
Confirm the Low Pulse Rate alarm activates.
6.
Restore the alarm limit.
S
P
O
2
1.
Set the High SpO
2
alarm limit below the measured SpO
2
value.
2.
Confirm that the High SpO
2
alarm activates.
3.
Restore the alarm limit.
4.
Set the Low SpO
2
alarm limit above the measured SpO
2
value.
5.
Confirm the Low SpO
2
alarm activates.
6.
Restore the alarm limit.
Alarm and Event Log
The Alarm and Event Log is a record of all events related to the device and to therapy. The log shows each event, when it occurred, and a brief
description. Information is retained even when you shut off the device or when power is lost. The log stores the most recent 6-months of
information with the exception of the event log, which stores the most recent 10,000 records. Older records are overwritten.
To access the Alarm and Event Log:
1.
In the menu bar, tap the
Options
icon.
2.
In the
Options
window, tap
Alarm & Event Log
.