Instructions for Use | Trilogy Evo Universal Accessories
Patient Monitors
79
To replace the sensor:
Tools required
:
-
T20 Torx screwdriver
-
Scratch awl or similar slender, pointed tool
To remove the FiO
2
sensor:
1.
Using a T20 Torx screwdriver, unscrew the FiO
2
sensor cover and set
aside. Note that the cover is also the FiO
2
sensor driver.
2.
Use the awl to pull the tab lock away from the plug to the unlock
position.
3.
While holding the tab in the unlock position, pull the white connector
straight out. Do not use the wires to pull the connector.
4.
Place the FiO
2
driver on the sensor and twist counterclockwise to
remove the sensor from the device.
5.
Because this sensor contains lead and a corrosive solution, contact a
licensed professional waste disposal service to dispose of this material.
To install a new FiO
2
sensor
:
1.
Place the sensor in the FiO
2
socket.
2.
Place the FiO
2
driver on the sensor and twist
clockwise to secure the sensor in the receptacle.
3.
Push the connector onto the pins on the sensor.
Ensure you have aligned the pins with the
receptacles on the connector.
4.
Replace the FiO
2
sensor cover and secure with
the screws.