Environmental
Operating
Temperature: 0°C to 40°C
Relative humidity: 5% to 90% RH, non-condensing
Atmospheric pressure: 62 to 106 kPa
Altitude: -1261 to 12,971 feet
Battery charging temperature: 5°C to 40°C
Transient operating temperature,
excluding high pressure oxygen
blending
-20°C to 50°C
Storage temperature
-25°C to 70°C
5% to 93% RH, non-condensing
Physical
Weight
6.3 Kg (13.9 lbs)
Size
19.3 cm D x 28.6 cm W x 24.5 cm H
7.6” D x 11.25” W x 9.65” H
Screen dimensions
8”, 20.32 cm
Ingress protection
IP22: protection against finger-sized objects and protected against
dripping water when tilted up to 15 degrees.
IEC 60601-1 classification
Type of protection against electric shock: Class II equipment
Degree of protection against electric shock – type BF applied part
Composition
This device does not contain natural latex rubber or dry natural rubber.
Expected service life
10 years
Internal memory capacity
2Gb
Mode of operation
Continuous
Maximum limited pressure
90 cmH
2
O
Electrical
AC input voltage
100V – 240V - 50/60 Hz 1.7A
DC input voltage
12/24V 6.5A
Internal and detachable Li-ion
batteries
7.5 hours nominal total run time per method in IEC 80601-2-
72 (each
battery)
Charge time for detachable and internal battery
from 0% to 80%: 2.5 hours
from 0% to 100%: 3.5 hours
Degree of protection against electric
shock
Type BF Applied Part
Audio
Alarm sound pressure level
Lowest setting: ≥
53.9 dBA
Highest setting: ≥
85.5 dBA
Sound pressure level - ventilator,
circuit and exhalation device
43.7 dBA as measured according to ISO 80601-2-12
Sound power level -ventilator, circuit
and exhalation device
51.6 dBA as measured according to ISO 80601-2-12
Oxygen
Low flow
0 to 30 l/min; maximum 10 psi
High pressure
280 to 60
0 kPa (41 to 87 psi)
Ventilator response to a 21 -90%
increase in oxygen concentration
< 30 seconds
Control Accuracy
Pressure
±(2 cmH
2
O + 4% of the setting)
Tidal volume
±(4 ml + 15% of setting)
FiO
2
±5 % FiO
2