Instructions for Use | Technical Specifications
EMC Information
95
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
IEC 61000-4-8
Electromagnetic
compatibility
(EMC). Part 4
-8:
Testing and
measurement
techniques. Power
frequency
(50/60
Hz)
magnetic field
immunity test
30 A/m
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical hospital
or home environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it
is used in such an environment.
Immunity Test
IEC 60601 Test Level Compliance Level
Electromagnetic Environment - Guidance
Conducted RF IEC 61000-
4-6
Electromagnetic
compatibility (EMC). Part
4-6: Testing and
measurement techniques.
Immunity to conducted
disturbances, induced by
radio-frequency fields.
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF communications
equipment should be used no closer to any part of
the device, including cables, than the
recommended 30 cm separation distance.
6 Vrms
Amateur Radio &
ISM Bands between
150 kHz and 80 MHz
6 Vrms
Amateur Radio & ISM
Bands between 150 kHz
and 80 MHz
Radiated RF
IEC 61000-4-3
Electromagnetic
compatibility (EMC)
- Part
4-3: Testing and
measurement techniques
- Radiated, radio-
frequency,
electromagnetic field
immunity test
10 V/m
80 MHz to 2.7 GHz
10 V/m
Up to 28V/m in
telecommunication
bands as specified in
clause 8.10 of IEC
60601-1-2
Up to 28V/m in
telecommunication bands
as specified in clause 8.10
of IEC 60601-1-2