Additional information
MC11 control unit
For the models with IPX4 protection or greater, in conformity with:
- EN 60601-2-52, IEC 60601-2-52, (special requirements for the safety, including essential perfor-
mance characteristics, of medical beds)
The following standards have been applied:
IEC/EN 60601-2-52, Section 201.6.2
Protection against electrical shock: Protection class: II
IEC/EN 60601-2-52, Section 201.7.6.3 Control panel symbols (depending on model, customer
requirements)
IEC/EN 60601-2-52, Section
201.8.11.3.2
EPR power supply cable: only for ≥ 2.5 m length
IEC/EN 60601-2-52, Section
201.9.2.2.5
Continuous operations: Control keypad only with keys
IEC/EN 60601-2-52,
Section 201.9.2.3.1
Unintentional movement: Prevented by means of a lo-
cking mechanism (such as Control box, Supervisor,
IPROXX
®
SE, IPROXX
®
, or Meditouch)
IEC/EN 60601-2-52, Section
201.9.6.2.1
Noise level: ≤ 65 dB(A)
IEC/EN 60601-2-52, Section
201.11.1.1
Temperature
IEC/EN 60601-2-52,
Section 201.11.6.5.101
Protection against water i
ngress: only for ≥ IPX4
IEC/EN 60601-2-52, Section 201.11.8 Loss of power (e.g. using a battery rechargeable bat-
tery), depending on customer requirements
IEC/EN 60601-2-52, Section
201.13.1.4
Special mechanical risks, depending on customer re-
quirements:
Prevented by means of a locking mechanism (such as
Control box, Supervisor, IPROXX
®
SE, IPROXX
®
, or
Meditouch)
IEC/EN 60601-2-52,
Section 201.15.3.4.1
Mechanical strength – handset (e.g. IPROXX
®
)
IEC/EN 60601-2-52, Section
201.15.4.4
Displays: Ready indicator is not required
IEC/EN 60601-2-52,
Section BB.3.3.3
Dimensions (depending on version and customer re-
quirements) – handset (e.g. IPROXX
®
)
IEC/EN 60601-2-52,
Section BB.3.4.1
Operational forces – handset (e.g. IPROXX
®
)