10
. Running a Control Solution Test
1.Insert a strip into the meter, and
you will hear: “
Hello, the meter is
ready.
” A full LCD screen will be
displayed briefly and then a blood
drop icon appears, then you will
hear: “
Apply blood now.
”
2.Press the “ ” key, and the meter will display
a letter “
C
” on the upper right corner of screen,
indicating the control test mode, then you will hear:
“
Apply control solution now.
” The test result will not
be stored in the memory.
C
(Please see Figure 10-1.)
C
3.Squeeze a small amount of
control solution on a flat surface
and apply to the front edge of the
test strip, then you will hear:
“
Testing in process
”.
4.Wait for 6 seconds, and the meter
will display the result, then you will
hear: “
The glucose level is
number/unit.
”
5. Remove the used strip and the
meter will power off with a display
of “
OFF
” on the screen, then you will hear: “
Goodbye.
”
(Please see
Figure 10-2.)
(Please see Figure 10-3.)
OFF
Figure 10-2
IMPORTANT!!!
Always make sure you press C key for control solution test,
otherwise the control solution test result will be stored in the
memory and affect your “AVG” results.
Figure 10-1
9-90
Figure 10-3