1
6
. Troubleshooting
Following is a summary of all Error Messages. These
messages help to identify certain problems, but do not
appear in all cases when a problem has occurred.
Improper use may cause an inaccurate result without
producing an error message or a symbol. In the event
of a problem, refer to the information in the table
under Solution.
Error messages:
E_1
:
The meter temperature is below its operating range.
(Audible message: “
The temperature is too low.
”)
E_2
:
The meter temperature is above its operating range.
(Audible message: “
The temperature is too high.
”)
E_3
:
Battery Low. No further tests can be performed.
(Audible message: “
Battery low.
”)
E_4
:
The blood sample is not sufficient (Audible message:
“
No sufficient blood.
”)
E_5
:
The strip is wet or used. (Audible message:
“
Strip is wet or used.
”)
E_6
:
The meter is not working properly, or the test strip is
damaged.
( Audible message: “
Error in meter or strip.
”)
E_7
:
Memory damaged.
( Audible message: “
Memory
damaged.
”)
Other error messages:
H I
:
Test result is higher than 600 mg/dL.
( Audible message: “
The glucose level is too high.
”)
LO
:
Test result is lower than 20mg/dL.
( Audible message: “
The glucose level is too low.
”)
E_1
E_2
E_3
E_4
E_5
E_6
E_7
H I
LO
31