Important Control Solution Test Information:
1.You must push the C key to distinguish the control
solution test from the blood glucose test. Do not
perform the blood test in Control Solution mode.
(If you perform the blood test in Control Solution
mode, and the test result will not be memoried.)
2. Use only Smartest Control Solution with Smartest
Glucose Meter.
3.Smartest control solutions are used to check that
the meter and the test strips are working together
as a system and that you are performing the test
correctly.
4.Smartest control solution contain a known amount
of glucose that reacts with Test strips.
The Medium and High level Control Solution are
intended to check the Monitoring system in
different measurement ranges.
5.Control Solution, meter, and test strips should
come to room temperature (20~25
℃
or 68~77
℉
)
before testing.
6.Shake the vial, discard the first drop of control
solution, and wipe off the Dispenser tip to ensure
a good sample and an accurate result.
7. Use only for 90 days after first opening. Record
the discard date (date opened plus 90 days)on the
control solution vial. Discard after 90 days.