X.OE-X
Y.OE-Y
zzzz
HH
Mode control
commands
Setpoint control
command
The output format is
X.OE-XbY.OE-YbZZZZbHH
where:
is parameter 1
is parameter 2
is parameter 3
is parameter 4, the bakeout timer; only appears for the command RS3.
Mode control commands are in the format MX
where X is a digit between 1 and 9 (Table 3-8).
Table 3-8.
Mode commands.
Mode Command
Function
1
Standby
2
Operate
3
Display voltage ( on LEDs)
4
Display current
5
Display pressure
6
Bakeout enable
7
Bakeout disable
8
Keyboard on
9
Keyboard off
These functions draw a close parallel to keyboard commands except
on/off commands. If the keyboard has been turned off from the
computer interface, pressing a key momentarily displays
Err 13.
The setpoint control command format is
SNPXXYY
where:
N
p
xx
Example 1:
is the setpoint number ( 1, 2 or 3)
is the parameter number (1, 2, 3 or 4)
are the two digits of the parameter, or if the parameter
requires four digits,
XXYY.
The control parameter of setpoint 1 is set to 0010. The command is
S130010.
Example 2:
The bakeout timer is set to 24 hours. The command is S3424.
Example 3:
Parameter 1 of setpoint 2 is set to 6.0 x 1
o -
9
•
The command is S2169.
Operation
3-23