8
TrueCPR Device Instructions for Use
Using the TrueCPR Device
Screen Illustrations
Calibration Screen
When you turn on the TrueCPR device, the calibration screen appears.
Compression Indicators
During compressions, the gray fan moves between the recoil zone and compression zone as shown.
Incorrect Compression Indicators
Calibration in progress.
• Do not apply pressure to
chest pad.
• Ensure chest pad and back
pad are separated during
calibration.
Calibration complete.
Begin chest compressions.
Compression in progress:
Gray fan representing
compression depth moves
across screen.
Sufficient depth: Dark green
wedge in compression
target zone.
Sufficient recoil: Dark green
wedge in recoil target zone.
Compression too shallow:
Arrow appears.
Compression > 6 cm:
Orange wedge appears
outside compression
target zone.
Recoil incomplete:
Arrow appears.