Data Management
Data Management
The TrueCPR device can store compression data for three 60-minute sessions, or up to six sessions
totaling 180 minutes. When all available memory has been used, the data from the oldest use is
overwritten automatically.
Data can be transferred to a computer with TrueCPR device-compatible software via USB connection.
Event reports can be printed directly from the software. For information on TrueCPR device reports and
full event review functionality, contact your local Physio-Control representative.
To initiate data transfer:
1 Connect a USB cable between the TrueCPR device and a computer with compatible software.
Note:
Use only a USB cable that meets the following specifications: USB 2.0 A-Male to Mini-B 5-pin
Male, 28/24 AWG cable with ferrite core (gold plated), 1 m (3 ft).
2 Turn on the TrueCPR device. If the software on the computer does not start automatically, open the
program manually.
3 Follow the instructions in the software on the computer to select the TrueCPR device and initiate data
transfer. Data transfer may take up to 5 minutes if memory is full. The following screens may appear on
the TrueCPR device:
WARNINGS
SHOCK HAZARD:
•
To avoid potential shock hazard, any device connected to the TrueCPR device through the USB
connector should be battery powered or certified to IEC 60601-1.
•
Do not use the USB connector if the TrueCPR device is damaged.
POSSIBLE ELECTRICAL INTERFERENCE:
Using cables or accessories that are not specified for use with the TrueCPR device may result in
increased emissions or decreased immunity from electromagnetic or radio frequency interference (RFI)
for this device. This could affect the performance of the TrueCPR device or equipment in close
proximity. Use only cables and accessories specified in these instructions for use.
Data transfer in progress.
Data transfer successful.
Data transfer failed.