3.2 Preparations and Start-Up
How to Start up the Instrument
(1)
Press the
Data Selector <2>
for one second. You will hear an acoustic signal and the
Display <1>
is
illuminated. The message
No CARD available in the unit! Please insert CARD.
is displayed.
(2)
Insert your card with the chip side facing down into the
Card Reader <4>
. The welcome screen is
displayed.
Note
If the card is defective or incompatible, a respective error message will be displayed.
(3)
Click the
symbol. The indications menu is displayed.
Indications Menu
You can now select an indication and begin treatment (refer to
on
page 15).
How to Switch off the Instrument
(1)
Press the
Data Selector <2>
for five seconds. The
Display <1>
goes out.
(2)
Remove the therapy card and unplug the cables from the
Ports <3>
.
3.3 Function Check
A function check is always necessary when you are unsure about whether the instrument is working prop-
erly.
How to Check the Function of the Instrument
(1)
Press the
Data Selector <2>
for several seconds until the instrument switches off.
(2)
Press the
Data Selector <2>
again until the instrument switches on:
•
If the instrument is properly working, the available options are displayed on the
Display <1>
.
You can operate the instrument without any further precaution.
•
If an error code is displayed, there is an instrument error. Proceed as described under
on page 14 .
PHYSIOMED ELEKTROMEDIZIN AG
3 Instrument Operation
DEEP OSCILLATION
®
Personal
11