Indications
Menu
IONOSON-IF-Expert
39
Furthermore, the treatment parameters will be shown. The illustration shows you to
which area of the body you can apply ultrasound. This illustration, however, is only a
suggestion.
(3)
Select the function
Therapy
.
(4)
If necessary, change the treatment parameters.
(5)
Increase the ultrasound dose with the
intensity control circuit II/dose <8>
.
Therapy time elapses.
(6)
After treatment, an acoustic signal will be issued. Turn the intensity down to 0.
Automatic Coupling Confirmation
The optical coupling confirmation at both transducer and instrument enables a sta-
ble and efficient therapy. If the ultrasonic waves are badly absorbed by the tissue of
the patient, the light at the transducer will start flashing. Now you can choose be-
tween counting the effective therapy time or running the timer independently from
the coupling
(setup menu).
Warning
Make sure you constantly keep moving the transducer on the skin of
the patient (dynamic treatment)! Avoid using the transducer at one
spot for a longer time (static treatment); this form of treatment might
lead to damages on the patient’s tissue!