Controls and Indicators
IONOSON-IF-Expert
7
2.1.12
Patient Lead Connector <12>
The
patient lead connector <12>
serves to plug in the patient lead. You can plug
plate electrodes, adhesive electrodes or other types of electrodes into the patient
lead.
Patient Lead - Mode(s) of Attachment
The colours of the connectors facilitate easy and cor-
rect attachment of the electrodes to the two circuits
and to their polarity.
Colour of the inner circle:
red = circuit I, blue = circuit II
Colour of the outer circle:
red = anode (+), blue = cathode (-)
Type BF component, not connected to protective ground wire.
Warning
Take care that the electrodes are by no means touched after the cur-
rent is turned up!
2.1.13
Ultrasonic Transducer Connector <13>
The
ultrasonic transducer
is connected to the instrument by means of the
ultra-
sonic transducer connector <13>.
The transducer is automatically calibrated after
it has been connected or after switching on the instrument.
Replacing the ultrasonic transducer during treatment will not affect the selected
mode of treatment. The dose, however, will automatically be turned down to 0 and
has to be increased again when continuing the treatment.
If no transducer has been plugged in, the message
Plug in head
will appear in the
lower status bar<4>
.
12
13
anode (+)
circuit I
anode (+)
circuit II
cathode (-)
circuit I
cathode (-)
circuit II