Simultaneous
Therapy
IONOSON-IF-Expert
35
6
Simultaneous Therapy
6.1
General Information and Preparations
Simultaneous therapy means the simultaneous treatment with ultrasound and
stimulation current. The ultrasonic transducer is one pole of the stimulation current
circuit, a plate electrode the other one.
It is advisable to use a plate electrode of minimum size EF50 as neutral electrode.
The neutral electrode is plugged into the socket No.1 (anode of circuit I) of the pa-
tient lead (page 7).
Make sure that there is enough space between the plate electrode and the trans-
ducer. If there is too little room, an error might be issued.
To ensure a stable contact for both stimulation and ultrasound therapy you must
use an (electrically) conductive ultrasound gel.
6.2
Therapy
Process of Treatment
(1)
Select
SONO
from the
display <3>.
(2)
Attach the neutral electrode (electrode #1) near the area of treatment.
(3)
Apply the ultrasound gel to the area of treatment.
(4)
Set the therapy time.
(5)
Select the treatment mode with the function
MODE
.
(6)
Select the treatment frequency.
(7)
If necessary, activate the acoustic coupling confirmation.
(8)
Select the desired current mode in the left half of the
display <3>
.
(9)
Set the parameters of the current mode (
from