Short Wave Therapy
14
PHYSIOTHERM-S
(2)
Check the parameters for compliance with the desired type of therapy.
(3)
Select the output power parameter with the
data selector <3>
.
(4)
By turning the
data selector <3>
,
slowly
increase the output power. Observe
the matching display and the well-being of the patient (verbal response).
(5)
Adjust the parameter setting during the treatment, if necessary.
Note
During active treatment it is not possible to switch directly from
continuous mode (CW mode) to pulsed mode and vice versa.
(6)
Press the
stop button <4>
to interrupt the treatment or wait until the
treatment duration has elapsed (signal tone).
(7)
Remove the electrodes and ask the patient how he feels.