5 Installation
E-836.1G Piezo Amplifier
PZ250E
Version: 1.0.1
13
In this Chapter
General Notes on Installation ...................................................................................... 13
Connecting the E-836.1G to the Protective Earth Conductor ..................................... 13
Connecting the Power Supply to the E-836.1G........................................................... 14
Connecting the Stage .................................................................................................. 15
Connecting the Control Input Voltage ......................................................................... 15
5.1 General Notes on Installation
Install the E-836.1G near the power source so that the power plug can be
quickly and easily disconnected from the mains.
Only use cables and connections that meet local safety regulations.
5.2 Connecting the E-836.1G to the Protective Earth
Conductor
INFORMATION
Observe the applicable standards for mounting the protective earth conductor.
Prerequisite
You have read and understood the general notes on installation (p. 13).
The E-836.1G is switched off, i. e. the power supply is
not
connected to the
power socket via the power cord.
Tools and accessories
Suitable protective earth conductor:
−
Cable cross-
section ≥0.75 mm
2
−
Contact resistance < 0.1 ohm at 25 A at all connection points relevant for
mounting the protective earth conductor
5