7 Maintenance
P54x Nanopositioner
PZ244E
Version: 1.2.0
29
In this Chapter
General Notes on Maintenance ................................................................................................... 29
Cleaning the P-54x ....................................................................................................................... 29
7.1
General Notes on Maintenance
NOTICE
Misalignment due to loosening screws!
The P-54x is maintenance-free and achieves its positioning accuracy as a result of the optimal
alignment of mechanical components and piezo actuators. Loosened screws cause a loss in
positioning accuracy.
Loosen screws only when instructed in this manual.
Do
not
open the P-54x.
7.2
Cleaning the P54x
NOTICE
Damage from ultrasonic cleaning!
Ultrasonic cleaning can damage the P-54x.
Do
not
do any ultrasonic cleaning.
Requirements
You have discharged the piezo actuators of the P-54x (p. 28).
You have disconnected the P-54x from the electronics.
Cleaning the P54x
Clean the surfaces of the P-54x with a cloth dampened with a mild cleanser or
disinfectant (e.g., isopropyl alcohol).
7
Maintenance