5 Installation
18
Version: 1.1.0
PZ272E
P-545 Nanopositioners
NOTICE
Damage due to unsuitable cables!
Unsuitable cables can damage the P-545 and the electronics.
Use cables provided by PI only to connect the P-545 to the electronics.
NOTICE
Damage due to improper mounting!
Improper mounting of the P-545 or incorrectly mounted parts can damage the P-545.
Only use the holes or threads intended for the purpose of fixing the P-545 and loads.
Install the P-545 so that the platform and all parts attached to it can move freely within the
entire travel range.
NOTICE
Damage due to incorrectly tightened screws!
Incorrectly tightened screws can cause damage.
Pay attention to the torque range (p. 46) specified for the screws used during installation.
INFORMATION
Extension cables can reduce the positioning accuracy of the P-545 or affect sensor processing
by the electronics.
Do
not
use extension cables. If you need longer cables, contact our customer service
department (p. 35).
5.2
Connecting the P-545 to the Protective Earth Conductor
INFORMATION
Pay attention to the applicable standards for connecting the protective earth conductor.
INFORMATION
If there is any vibration in your application, secure the screw connection for the protective
earth conductor in a suitable manner (e.g., with liquid adhesive) to prevent it from
unscrewing by itself.