6 Starting and Operating
P-545 Nanopositioners
PZ272E
Version: 1.1.0
29
6.3
Discharging the P-545
The P-545 must be discharged in the following cases:
Before Installation
When the P-545 is not used but the electronics remain switched on to ensure
temperature stability
Before demounting (e.g., before cleaning and transporting the P-545 and for
modifications)
The P-545 is discharged via the discharge resistor inside the electronics from PI.
Discharging a positioner connected to the electronics
In closed-loop operation:
1.
Switch off the servo mode on the controller.
2.
Set the piezo voltage to 0 V on the controller.
In open-loop operation:
Set the piezo voltage to 0 V on the electronics.
Discharging a positioner not connected to the electronics
Connect the positioner to the switched-off electronics from PI.