6 Starting and Operating
P-63x Nanopositioner
PZ247E
Version: 1.1.0
25
Operating the P-63x
Follow the instructions for starting and operating the P-63x in the manual for the
electronics (p. 11) used.
6.3
Discharging the P-63x
The P-63x must be discharged in the following cases:
Before Installation
When the P-63x is not used but the electronics remain switched on to ensure
temperature stability
Before demounting (e.g., before cleaning and transporting the P-63x and for
modifications)
The P-63x is discharged via the discharge resistor inside the electronics from PI.
Discharging a positioner connected to the electronics
In closed-loop operation:
1.
Switch off the servo mode on the controller.
2.
Set the piezo voltage to 0 V on the controller.
In open-loop operation:
Set the piezo voltage to 0 V on the electronics.
Discharging a positioner not connected to the electronics
Connect the positioner to the switched-off electronics from PI.