10 Technical Data
Q-522 Miniature Stage
MP134E
Version: 1.0.1
49
10.1.2 Maximum Ratings
The Q-522 stage is designed for the following operating data:
Maximum Operating
Voltage
Maximum Operating
Frequency
Maximum Power
Consumption
48 V
20 kHz
10 W
10.1.3 Ambient Conditions and Classifications
The following ambient conditions and classifications must be observed for the Q-522:
Area of application
For indoor use only
Maximum altitude
2000 m
Air pressure
Q-522.xy0:
1100 hPa to 0.1 hPa
Q-522.xyU:
1100 hPa to 10
-9
hPa
Relative humidity
Highest relative humidity 80 % for temperatures up to 31 °C,
non-condensing
Decreasing linearly to 50 % relative humidity at 40 °C,
non-condensing
Operating temperature
0 °C to 40 °C
Storage temperature
-20 °C to 70 °C
Transport temperature
-20 °C to 70 °C
Maximum bakeout temperature
for vacuum-compatible products
Q-522.xyU:
120 °C, for 12 hours, only in
switched-off
state
Overvoltage category (in acc.
with EN 60664-1:2007 /
VDE 0110-1)
II
Protection class
(acc. to EN 61140 /
VDE 0140-1)
I
Degree of pollution
(acc. to EN 60664-1:2007 /
VDE 0110-1)
1
Degree of protection
(acc. to IEC 60529)
IP20