9
3.
Obtaining a blood sample (i)
Use the pre-set lancing device to puncture the desired site. The size of the drop should be at least 0.5 microliter
(μL) of volume for glucose test. Gently squeeze the punctured area to obtain another drop of blood. Be careful
NOT
to smear the blood sample.
4.
Apply the sample (j)
Gently apply the drop of blood to the absorbent hole of the test strip at a tilted angle.
Confirmation window should be completely filled if enough blood sample has been
applied. Do
NOT
remove your finger until you hear a beep sound.
If the blood volume is insufficient, the triangle icon with exclamation mark will appear
on the meter display to remind you to refill the second drop of blood to the same test
strip within 3 seconds.
If the blood volume is still underfill after the second drop of blood applying, E-3 error
message will appear on the display with the triangle icon with exclamation mark.
NOTE:
• Do not press the punctured site against the test strip or try to smear the blood.
• If you do not apply a blood sample to the test strip within 3 minutes, the meter will automatically turn off. You
must remove and reinsert the test strip to start a new test.
• The confirmation window should be filled with blood before the meter begins to count down. If there is
insufficient blood volume in the test strip confirmation window, you can apply second drop of blood to the
same strip before countdown begins. The meter will display an “E-3” error message if there is still underfill, in
which case
discard the used test strip and retest with a new one
.
• If you have trouble filling the confirmation window, please contact the consumers’ toll-free number for Italy
800 900 080 and for France 01.55.93.26.48. for assistance.
5.
Read Your Result
The result of your blood glucose test will appear after the meter counts
down to 0. For result readings please refer to paragraph “Measurement Result
Readings”. The blood glucose result will be stored in the memory automatically.
6.
Pull out the used test strip (k)
Pull out the used test strip. Use a sharp bin to dispose of used test strips. The
meter will switch itself off automatically.
WARNING:
The used lancet and test strip may be biohazardous. Please discard them carefully according to your local
regulations.