14
This product complies with the Directive 2012/19/EU
.
The crossed bin symbol on the appliance indicates that the product, at the end of its life, must be
disposed of separately from domestic waste, either by taking it to a separate waste disposal site for
electric and electronic appliances or by returning it to your dealer when you buy another similar
appliance. The user is responsible for taking the appliance to a special waste disposal site at the end
of its life. If the disused appliance is collected correctly as separate waste, it can be recycled, treated
and disposed of ecologically; this avoids a negative impact on both the environment and health,
and contributes towards the recycling of the product’s materials.
For further information regarding the waste disposal services available, contact your local waste disposal agency
or the shop where you bought the appliance.
Caring for Your Test Strips
Storage conditions for strips: 2ºC (35.6ºF) ~ 30ºC (86ºF) (Vial Pack); 2ºC (35.6ºF) ~ 30ºC (86ºF) (Individual Pack), 10 to
95% relative humidity (non-condensing).
• relative humidity (non-condensing). Do
NOT
freeze.
• Store your test strips in their original vial only. Do not transfer to another container.
• Store test strip packages in a cool dry place. Keep away from direct sunlight and heat.
• After removing a test strip from the vial, immediately close the vial cap tightly.
• Touch the test strip with clean and dry hands. Use each test strip immediately after removing it from the vial
or individual foil packet.
• Write the opening date on the vial label when you first opened it. Discard remaining test strips after 6 months.
• Do not use test strips beyond the expiration date. This may cause inaccurate results.
• Do not bend, cut, or alter a test strip in any way.
• Keep the strip vial away from children since the cap and the test strip may be a choking hazard. If swallowed,
promptly see a doctor for help.
For further information, please refer to the test strip package insert.
Caring for Your Lancing Device
Clean the outside of the finger-pricking device with lukewarm water or mild detergents and dry it carefully.
Hospital must follow their standard disinfection protocol.
Important Control Solution Information
• Use only Pic GlicSolution control solutions with your meter.
• Do not use the control solution beyond the expiration date or 3 months after first opening. Write the opening
date on the control solution vial and discard the remaining solution after 3 months.
• It is recommended that the control solution test be done at room temperature 20°C to 25°C (68°F to 77°F). Make
sure your control solution, meter, and test strips are at this specified temperature range before testing.
• Shake the vial before use, discard the first drop of control solution, and wipe off the dispenser tip to ensure a
pure sample and an accurate result.
• Store the control solution tightly closed at temperatures between 2°C to 30°C (35.6°F to 86.0°F). Do
NOT
freeze.
WARRANTY
The product is guaranteed 3 years against any conformity defect in normal conditions of use as provided for by the
instructions for use. The warranty shall not therefore apply in the case of damages caused by improper use, wear
or accidental events. For the duration of warranty on conformity defects please refer to the specific previsions of
national laws applicable in the country of purchase, where provided.