28
Floors should be wood, concrete or
ceramic. If the floors are covered with
synthetic material, the relative humid-
ity should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Manufacturer’s guide and declaration – Electromagnetic Immunity
The Thermal Inhaler TermalVap
02 038069 100 000 is intended for use in the electromagnetic environ-
ment specified below. The customer or user of the TermalVap
02 038069 100 000 must ensure that the
device is used in this environment.
Electromagnetic Environment - Guidance
Test level
Immunity test
+/-6kV contact
+/-8kV air
Electrostatic dis-
charge (ESD)
EN 61000-4-2
Level of compliance
The appliance does
not change its state
+/-2kV for power
supply
+/-1kV for signal
conductors
Electrical fast tran-
sient/burst
EN 61000-4-4
The appliance does
not change its state
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
+/-1 kV differential
mode
+/-2 kV common
mode
Surge
EN 61000-4-5
The appliance does
not change its state
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
If the TermalVap thermal inhaler user
requires the unit to be in continued
operation, we recommend powering it
from an uninterruptible power supply.
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40 % U
T
(60% dip
in U
T
) for 5 cycles
70 % U
T
(30% dip
in U
T
) for 25 cycles
<5 % U
T
(>95%
dip in U
T
) for 5 s
Voltage dips, short
interruptions and
voltage variations EN
61000-4-11
- - -
The mains frequency magnetic fields
must have levels characteristics of
installations in commercial or hospital
environments.
3A/m
Magnetic Field at
mains frequency
(50/60 Hz)
EN 61000-4-8
The appliance does
not change its state
Note: U
t
is the mains voltage