29
Portable and mobile RF communication
devices must not be
used closer to any part
of the TermalVap device,
including the cables,
than the separation distance
calculated by the equation
applicable to the transmitter
frequency. Recommended
separation distance
d = [3.5 / V
1
]
√
P
d = [12 / E
1
]
√
P
from 80 MHz to
800MHz
d = [23 / E
1
]
√
P
from 800 MHz to
2.5GHz
Where P is the maximum rated output
power of the transmitter in Watts (W)
according to the transmitter manu-
facturer and d is the recommended
separation distance in metres (m). The
intensities of the fields from fixed RF
transmitters, as determined in an elec-
tromagnetic study of the site
a)
,
may be less than the level of conformity
of each frequency interval
b)
.
Interference may occur near devices
marked by the following symbol:
Manufacturer’s guide and declaration – Electromagnetic Immunity
The Thermal Inhaler TermalVap
02 038069 100 000 is intended for use in the electromagnetic environ-
ment specified below. The customer or user of the TermalVap
02 038069 100 000 must ensure that the
device is used in this environment.
Electromagnetic Environment - Guid-
ance
Level indicated by
IEC 60601-1-2
Immunity test
3Vrms 150kHz to
80MHz
(for non life-
supporting
devices)
Conducted immunity
EN 61000-4-6
Level of compliance
V
1
= 3 V rms
3V/m 80MHz to
2.5GHz
(for non life-
supporting
devices)
Radiated immunity
EN 61000-4-3
E
1
= 3 V / m
Note 1: At 80 MHz and 800 MHz the range of the highest frequency is applied.
Note 2: These guidelines may not cover all situations. Electromagnetic radiation is affected by absorption and
reflection by structures, objects and people.
a) The field intensities for fixed transmitters like base stations for radio telephones (cellular and cordless) and ter-
restrial mobile radios, HAM radio devices, AM and FM radio transmitters and TV transmitters can not be theoreti-
cally foreseen accurately. In order to establish an electromagnetic environment caused by fixed RF transmitters
an electromagnetic study of the site must be considered. If the field intensity measured in the location where the
appliance is used exceeds the applicable level of conformity above the normal operation of the appliance itself
must be placed under observation.
If abnormal performance is noted additional measures may required such as different orientation or positioning
of the appliance.
b) The field intensity on a frequency range from 150 kHz to 80 MHz must be less than 3 V/m.