11
About the unit
4.4
Noise-free resolution and conversion time
The table below shows the number of
bits of resolution for the full range of
conversion times.
Conversion
Voltage range
time
& Noise-free resolution (bits)
per
channel
±
39 mV
±
78 mV
±
156 mV
±
313 mV
±
625 mV ±1250 mV
±
2500 mV
ADC-24 only
ADC-20 and ADC-24
660 ms
17
18
19
20
20
20
20
340 ms
17
18
19
19
19
20
20
180 ms
16
17
18
19
19
19
19
100 ms
16
17
18
18
18
19
19
60 ms
15
16
17
18
18
18
18
4.5
ADC-20/ADC-24 terminal board
For easy connection to the DB25 connector, we recommend that you use the
ADC-20/
ADC-24 terminal board
, part number PP310. This has screw terminals to allow you
to connect wires to all of the data logger's inputs and outputs without soldering. It also
has space for voltage-divider resistors, a temperature sensor and a quad op-amp.
See the
ADC-20/ADC-24 Terminal Board User's Guide
for instructions on connecting
signals to the terminal board. If you are not using a terminal board, you can still refer
to the Terminal Board User's Guide for information on connecting various types of
signal to the data logger.
http://www.elso.sk/product.php?id_product=2136
4.6
LED
The ADC-20 and ADC-24 have an LED, next to the entry point of the USB cable, that
you can use to verify that the unit is working. The LED flashes whenever the unit is
taking readings. It also flashes briefly during "enumeration", the process that PicoLog
uses to detect all Pico USB devices plugged in to the computer.
adc20.en r6
Copyright © 2005–2016 Pico Technology Ltd. All rights reserved.