162
162
162
162
L0
Set to the default identification string.
L1
Set to the compatible identification string.
SYSTEM:IDNSTR "Manufacturer,Product"
Change the multimeter’s identification string, such as the maker’s and the
product’s names (be sure to dimension a string variable with at most 39
characters).
DATA:POINts?
Query the number of readings stored in the multimeter’s internal memory.
*RST
Reset the multimeter to the power-on configuration. This command can not
clear the error queue.
*IDN?
Read the multimeter’s identification string (be sure to dimension a string
variable with at least 35 characters).
Other Interface Commands
SYSTem:LOCal
Place the multimeter in the
local
mode. All buttons on the front panel are fully
functional.
SYSTem:REMote
Place the multimeter in the
remote
mode. All buttons on the front panel, except
the LOCAL button, are disabled.
C.9 The SCPI Status Pattern
Status registers are provided in the same way by all SCPI equipment.
And there are three register groups with various equipment conditions
recorded by the status system. They are
the Status Byte Register
,
the