English
19
1. Display
2. Menu keys
3. Modification keys
The intensity and the
different parameters can
be changed during
stimulation and in
editing mode by
means of these keys.
4. On / Off – Button
5. Battery compartment
6. Output sockets
Operation of the MiniFG
Switch on the Device
Switch on the device by means of the key. The
MiniFG
will start with Program 1 (P1).
How do I carry out the stimulation?
Selection of the Program
You change to the next standard or user program by operating
the modification keys
(Figure 1). During this, the standard
programs 1 to 6 indicated with ‘P’ will run through initially. Then,
the user programs identified with ‘U’ will appear, as far as they exist.
When the final program has been reached, the device will spring
back to program 1 (P1) at the next keystroke.
The relevant operating parameters for each program will be shown
in the display:
Sinus
Rectangle
Positive – Negative
Positive
C
Continuing
P
Program (7-20-7-20-7min.)
F
6 Frequencies at 10min.
Figure 1
1
2
3
4
5
6