Pierenkemper Operating Manual - PierenPlus basic
50
Programme overview
No. Remark
Indications
Description
Frequency/
Hz
Pulse
width/µs
Timer/
min
1
Gate
Control 1
Acute, nociceptive
pain, acute
and chronic
neuropathic pain
Both channels equal
100
200
30
2
Gate
Control 2
Alternatively to
Programme 1
Both channels equal
80
150
30
3
Low
frequency
Chronic,
nociceptive pain,
improvement of
blood circulation,
Kaada-Application
Both channels equal
2
250
30
4
Low
Frequency
+ Gate
Control
Combined high
and low frequency
stimulation for
special application
area
Channel 1:100 Hz
Channel 2: 2 Hz
100
2
200
30
5
Adjust-
ment
Muscle-related
pain
First 10 minutes at 100
Hz, after that
20 minutes 2 Hz
100/2
150/200
10/20
6
HAN
Ideal for nearly all
pain
indications
3 seconds at 100 Hz,
after that
3 seconds at 2 Hz
100/2
150/200
30
7
Burst
Alternatively to
Progr. 3
0,25 seconds at 100 Hz,
after that
0,25 seconds pause (=>
2 Hz)
100
150
30
8
Modu-
lation
Alternatively to
other programmes
with insufficient
therapy success
Automatic frequency
order:
2 -> 80 -> 2 Hz in 15
seconds
2-80
200-100
30
9
Muscle
training
Atrophy
prophylaxis
Rising time 2 seconds,
working time 5 seconds,
falling time 1 seconds,
pause time 12 seconds;
automatic intensity
adjustment
50
250
30